1.
Biochem Med (Zagreb) ; 34(2): 020802, 2024 Jun 15.
Artigo em Inglês | MEDLINE | ID: mdl-38665873

RESUMO

We present two cases from the neonatal department with cerebrospinal fluid examination. We revealed a striking discrepancy in polymorphonuclear (PMN) and mononuclear (MN) cell counts using conventional light microscopy in comparison with automated analyzer Sysmex XN-1000 (PMNs - 13 vs. 173x106/L, MNs - 200 vs. 67x106/L in case 1 and PMNs - 13 vs. 372x106/L, MNs - 411 vs. 179x106/L in case 2). We revealed the dominant presence of hemosiderophages in both cases in cytospin slide. Even though Sysmex XN-1000 offers fast examination with a low sample volume, there is possibility of misdiagnosis, with negative impact on the patient.


Assuntos
Microscopia , Humanos , Recém-Nascido , Microscopia/métodos , Masculino , Feminino , Neutrófilos/citologia , Neutrófilos/patologia , Líquido Cefalorraquidiano/citologia , Contagem de Leucócitos , Leucócitos Mononucleares/patologia , Leucócitos Mononucleares/citologia